Principle of Testing NAWI
Every instrument should be accurate and correct to be permitted for use in any commercial transaction, particularly at the POS. An instrument may be accurate but may not be correct. To be accurate, the instrument must meet the applicable tolerance limit (MPE) and some metrological requirements. It should pass some specific performance tests. To be correct, an instrument, should not only be accurate, it should meet some specific legal requirements also - type (design) or model approval, appropriate class for usage, descriptive markings, proper maintenance etc.
Verification: Basics for Non-automatic Weighing Instruments
The Legal Metrology (General) Rules, 2011 Seventh Schedule Heading A Part II
Accuracy Class
All NAWI's are divided into 4 categories:
- Class I: Special accuracy - ultra micro, micro,
semi micro, macro
- Class II: High accuracy - precision balances
- Class III: Medium accuracy - NAWI's for trade
use
- Class IV Ordinary accuracy - For low accuracy trade use
Accuracy classes are determined in terms of Table 17
Table 17
| Class |
e value in g |
Verification Scale Interval n = max/e |
min capacity |
| I |
0.001 or above |
50 000 |
100e |
| II |
0.001 to 0.05 0.1 or above |
100 to 100 000 5 000 to 100 000 |
20e 50e |
| III |
0.1 to 2 5 or above |
100 to 10 000 500 to 10 000 |
20e 20e |
| IV |
5 or above |
100 to 1000 |
10e |
- Check if declared e value matches with the accuracy class
- Find n = max/e and check whether it matches with the accuracy class
- Check whether min value matches with the accuracy class
- Some instruments do not declare minimum capacity. Calculate the minimum capacity from the table and check weather the instrument can measure below that quantity
Preliminary Checking
Descriptive Plate [Clause 7(1)]
All machines should have a permanently fixed descriptive plate with markings:
- Manufacturer ´s Mark and Model no.
- Accuracy Class - I / II / III / IIII
- Verification Scale Interval e
- Scale Interval d [when d < e ]
- Maximum Capacity Max
- Minimum Capacity Min
An example
A NAWI declares max 30 kg having e=10 g, class III with min 200 g.
Here, n = max/e = 30,000/10 = 3,000. In Table 17, the n value of class III instruments lie between 100 to 10,000. Again, a class III instrument has min value of 20e = 20 X 10 = 200 g.
So, all conditions satisfy.
Equipment
- Model Approval Certificate
- Set of Working Standard Weights, (having error not more than 1/3rd of the MPE prescribed for the machine under test. Clause 9(2)) as per table below:
| Accuracy Class |
Working Standards Weights Required |
| Class I and II |
1 X 10 mg,
2 X 20 mg, 1 X 50 mg, 1 X 100 mg,
2 X 200 mg, 1 X 500 mg, 1 X 1 g, 2 X 2 g, 1 X 5 g, 1 X 10 g, 2 X 20 g, 1 X 50 g,
1 X 100 g, 2 X 200 g, 1 X 500 g, 1 X 1 kg, 2 X 2 kg and 1 X 5 kg.
|
Class III and IV up to 30 kg |
- Set of 50 mg to 5 kg
- 10 kg and 20 kg
|
Class III and IV above 30 kg up to 3 t |
- 1g to 5 kg
- 1 t
- Availability of 2 t further weight
|
Class III and IV above 3t |
- Set of 100 g to 5 kg
- 3 t
- Availability of 50% of the maximum capacity further weight
|
- Suitable substitution weights
Error Calculation
[Error E = I + ½e - L - Δl]
| Load L |
Indication I |
Addl. Load Δl |
Changed Indication |
Error |
| 500 g |
500 g |
0.3 g |
501 g |
+ 0.2 g |
Error E = 500 + 0.5 - 500 - 0.3 = + 0.2 g (considering e = 1 g)
- Additional load will start from 0.1e and gradually increased by 0.1 (1/10 e) till the indication rises by 1e.
Metrological requirements
- An instrument must be allowed to warm up for about half an hour testing.
- All NAWI's should pass following tests under normal conditions and in the sequence specified below.
- Once an instrument fails in a test, it is no use carrying out any further test.
Essential Tests
Seventh Schedule, Part II, Clause 8 (ii)
Tests shall be carried out to verify compliance with the following requirements:
- Value of maximum permissible error,
- Maximum permissible error for net and tare values,
- Scale interval of the tare weighing device,
- operation of the tare device shall be visible
- repeatability (3 weighing on classes III and IV and 6 weighing for I and II )
- Eccentric loading
- Discriminatlon
1. Error Limit
Maximum Permissible Error
All NAWI may have MPE as per table below.
Table 20
| |
Class I |
Class II |
Class III |
Class IV |
MPE ± |
| Load in e |
min to 50 000 |
min to 5 000 |
min to 500 |
min to 50 |
0.5e |
| Load in e |
50 001 to 200 000 |
5 001 to 20 000 |
501 to 2 000 |
51 to 200 |
1e |
| Load in e |
above 200 000 |
20 001 to 100 000 |
2 001 to 10 000 |
201 to 1 000 |
1.5e |
Checking Error Limit
Select at least three loads - one near minimum capacity, another two where the MPE changes over in Table 43. For example, 5e, 500e and 2000e for a class III instrument.
- Apply the loads, one by one and check change in indication in each case.
- If there is no change in indication, the instrument is PASSED and no further test is necessary.
- If the load and the indicated value are not same in the ½e range, the instrument FAILS.
- In other cases, further test is necessary.
When the load is in ± 1e range
- Apply the load and allow to show the stable indication of load + 1e.
- Then add additional 0.5e load.
- If the indication does not change, the instrument PASSES.
- If it changes and finally stabilizes at load + 2e, it FAILS
- When the indication stabilizes at load - 1e, add additional load of 0.5e.
- If the indication changes and become stable at load value, the instrument PASSES.
- If the indication does not change, it FAILS.
When the load value is in ± 1.5e range
- If the stable indication is ± 1e from the load value, the instrument PASSES.
- If the indication is more than that, it FAILS.
2. Repeatabilty
This test is designed to check whether the instrument gives consistent reading for the same load when the load is placed on the same position on the receptor for a number of times.
The difference between the maximum and minimum results of several weighings of the same load shall not be greater than the absolute value of the MPE of the instrument for that load.
[Clause 3(7)(1)]
For example, if the MPE for the load is equal to ±1e, the absolute value of this error is | ±1 | e = 1e.
Checking Repeatabilty
Determine the indicating type of the instrument
- non-self indicating like platform instrument or equal arm instrument
- analog indicating
- digital indicating
Test will be different for different instruments.
- Select a load just less than the second MPE change point. 1995e for Class III instrument, for example.
Non-self-indicating Instruments
Platform Instrument |
Equal Arm Instrument |
| 1. Set the instrument at zero. |
1. Set the instrument at zero. |
| 2. Apply the load to the load receptor. |
2. Apply the load to the goods receptor. |
| 3. Bring the indicating element to the equilibrium position using the proportional weights and/or steel yard poises and record the indication. |
3. Apply standard weights to the load receptor until the indicating element is in the equilibrium position and record the weight. |
| 4. Remove the load from the load receptor. |
4. Remove the loads from both the receptors. |
| 5. Reset the instrument to zero if the indication is not zero. |
5. Reset the instrument to zero if the indication is not zero. |
| 6. Repeat steps 2 to 5 twice more. |
6. Repeat steps 2 to 5 twice more. |
| 7. Check the difference between maximum and minimum indication does not exceed the absolute value of MPE for the applied load. If exceeds, it FAILS. |
7. Check the difference between maximum and minimum indication does not exceed the absolute value of MPE for the applied load. If exceeds, it FAILS. |
Analog Indicating Instruments
- Set the instrument to zero.
- Apply the load and record indication.
- Remove the load.
- Reset the instrument to zero if the indication is not zero.
- Repeat steps 2 to 4 twice more.
- Check the difference between maximum and minimum indication does not exceed the absolute value of MPE for the applied load.
Digital Indicating Instruments
- The instrument is set to 0 and then a load is applied.
- Add 0.5e load.
- Add several additional loads of 0.1e one by one until the reading changes and stabilize.
- Remove the 0.5e load. (The original load and the additional loads will remain)
- Note the reading (R1)
- Remove all the loads at a time and reset the indication to 0, if required.
- Again place the load and the additional loads at a time.
- Note the reading (R2) and remove the loads at a time.
- Repeat the procedure for third time and note the reading (R3)
SITUATION I: If R1 = R2 = R3, the instrument PASSES.
SITUATION II: If the reading R2 or R3 differs from R1 by more than ±1e, the instrument FAILS.
SITUATION III: If the reading R2 or R3 differs from R1 by ±1e, further tests are required.
- When the reading is R2, add additional loads of 0.1e till next changeover point is reached, that is R2 + e.
- Let total additional load be Dl.
- Now, calculate I2 = R2 + 1/2e - Dl.
- Repeat the same procedure for R3 and calculate I3.
- Rename R1 as I1.
- Find out which has the highest and which has the lowest value and calculate the difference
- If the difference is less than absolute value of the MPE for the load used, the instrument PASSES.
- If the difference is greater than absolute value of the MPE for the load used, the instrument FAILS.
3. Eccentricity
The indications for different positions of a load shall meet the MPEs, when the instrument is tested according to Clause 3(7)(ii)(a) and (b).
Determine whether the load receptor on the instrument has:
- four or less points of support;
- more than four points of support;
Instrument with a Load Receptor with Four or Less Points of Support
On an instrument with a load receptor with four or less points of support, a load corresponding to one-third of the sum of the maximum capacity and the corresponding maximum additive tare effect shall be applied. Large standard working weights should be used in preference to several small standard weights. Apply the load centrally in the segment as shown in the figure.
- Zero the instrument.
- Apply one-third Max plus maximum additive tare (if applicable) at one load position.
- Record the load and the indication.
- Determine if the indication is within the MPE for the load applied.
- Remove the load
- Repeat steps 1 to 5 for each segment.
Instrument with a Load Receptor with more than Four Points of Support
( Weigh-bridge, for example)
- Determine the number of support points.
- Divide the load receptor into n approximately equal segments, where n is the number of points of support.
- Zero the instrument.
- Apply 1/(n - 1) Max plus Max additive tare (if applicable) at the first load position as shown in the diagram.
- Record the load and the indication.
- Determine if the indication is within the MPE for the load applied.
- Remove the load.
- Repeat steps 3 to 7 at all the other locations in turn for all points of support.
- Determine if the indication is within the MPE for the load applied at every position.
4. Zero-setting
After zero setting the effect of zero deviation on the result of the weighing shall not be more than 0.25e; however, on an instrument with auxiliary indicating devices this effect shall be not more than 0.5d (Clause 4.5.ii).
If the instrument being tested has an auxiliary indicating device, activate the zero-setting device as shown below. If the indication is reading anything other than zero, the instrument does not meet the requirements of ±0.5d and has failed.
Determine, whether the instrument is
- non-self-indicating
- analog indicating; or
- digital indicating
Non-self Indicating Instrument
During any test,when the entire load has been removed check visually that the instrument has returned to within ±0.25e of its equilibrium position. If it has not, apply 0.25e on the appropriate load receptor. Then:
- If the indicator has moved through the equilibrium point the instrument PASSES; or
- if the indicator has not moved through the equilibrium point the instrument FAILS.
Analog Indicating Instrument
The accuracy of the zero-setting device of an instrument with analogue indication can be checked at any stage during the testing of the instrument. It is essential that an instrument is set on zero before starting any test. After the completion of one of the test sequences check visually that the instrument has returned to within ±0.25e of zero.
Digital Indicating Instrument
Determine whether the instrument has non-automatic or semi-automatic or automatic zero-setting by conducting a simple test.
- (a) Apply a load of approximately 5e.
- (b) Zero the instrument and then remove the load.
- (c) Wait for the automatic zero setting to occur and the indication displays zero.
If the display does not return to zero after 15 seconds, the instrument does not have automatic zero setting and it should be tested as non or semi-automatic zero-setting instrument.
Non-automatic/Semi-automatic Instrument
- Activate the zero-setting device.
(a) Load the instrument using a standard weight that is within the zero-setting range (0 to 4% of Max, normally ±2% around zero).
(b) Add additional standard weights to take the total load just below the next changeover point.
- Re-set the indication to zero using the zero-setting device.
- Apply 10e to the load receptor.
- Apply an additional 0.25e. If the indication:
(i) remains unchanged go to step 5;
(ii) changes and stabilizes at +1e from the original indication: FAIL
- If the indication remains unchanged in step 4, apply an additional 0.5e. If the indication:
(i) changes and stabilizes at +1e from the original indication: PASS
(ii) remains unchanged: FAIL
Automatic zero Indicating Instrument
- (a) Apply a load of approximately 5e.
(b) Zero the instrument and then remove the load.
(c) Wait for the automatic zero setting to occur and the indication displays zero, this should take a minimum of 5 seconds.
- Quickly apply 10e to the load receptor.
- Apply an additional 0.25e. If the indication:
(i) remains unchanged go to step 4;
(ii) changes and stabilizes at +1e from the original indication: FAIL
- If the indication remains unchanged in Step 3, apply an additional 0.5e. If the indication:
(i) changes and stabilizes at +1e from the original indication: PASS
(ii) remains unchanged: FAIL
5. Weighing Performance
This test procedure is used to establish the weighing performance of the instrument at several loads. When loading and unloading weights, the load shall be progressively increased and decreased. The loads shall be applied evenly distributed over the platform.
When loading and unloading the instrument it must not be allowed to zero track. This is achieved by maintaining a suitable load on the instrument when loading and unloading.
- Select at least five different loads so that following loads may be covered by adding them suitably in increasing order - Minimum capacity, 1st MPE changeover point, 2nd MPE changeover point, 1/2 maximum and maximum capacity
- Zero the instrument.
- Apply the loads one by one increasing from minimum to maximum.
- Note, if the indication is within the MPE after each load is added.
- When the maximum weight is reached, an additional weight of 10e is added to see if the over-range blinking is set correctly.
- Remove the load in decreasing order and see if the indication is within the MPE after a load is removed.
- After the last load is removed check if the indication has returned to within ±0.25e of zero.
Instruments with a maximum capacity greater than 3 tonne
When testing instruments with a maximum capacity greater than 3 tonne, instead of standard weights any other constant load made up of substitution material may be used, provided that for instruments with maximum capacity:
- 3 t < Max ≤ 15 t standard weights to at least 3 tonne, or the maximum capacity which ever is the lesser, plus any additive tare are used, and
- > 15 t, standard weights to at least 20% maximum capacity plus any additive tare are used.
6. Discrimination
This test is done to check if the instrument is capable of sensing a small change in load.
The actual scale interval for a class 1 or 2 digital instrument may be d and not e. As the majority of instruments to be verified will be class 3 or 4 where e = d, e is used in the procedure. If the instrument to be tested has d ≠ e then e becomes d in this instance for the whole procedure.
Determine whether the instrument is:
- non-self-indicating;
- analogue indicating; or
- digital indicating
Non-self-indicating Instrument
An extra load equivalent to 0.4 times the absolute value of the MPE for the applied load when gently placed on or withdrawn from the instrument at equilibrium shall produce a visible movement of the indicating element [Clause 3(8)(i)].
Indicator Displacement Table
| Visible movement |
Accuracy Class |
| 1 mm |
Class I or II |
| 2 mm |
Class III or IV (with max up to 30 kg) |
| 5 mm |
Class III or IV (with max above30 kg) |
- Zero the instrument.
- Apply the load to the load receptor.
- Bring the instrument to its equilibrium position.
- Gently apply an extra load of 0.4 times the absolute value of the MPE for the applied load on the load receptor.
- Observe if there is a visible amount of movement of the indicator.
- Remove the load.
- Determine whether the instrument has passed or failed.
Analogue Indicating Instrument
An extra load equivalent to the absolute value of the MPE for the applied load when gently placed on or withdrawn from the instrument at equilibrium shall cause a permanent displacement of the indicating element corresponding to not less than 0.7 times the extra load [Clause 3(8)(ii)(a)]
1. Zero the instrument.
2. Apply a load to the load receptor and, bring the indication to a mark by applying a small amount of extra material to the load receptor.
3. Record the initial indication as I1.
4. Gently apply an extra load equal to the absolute value of the MPE for the applied load on the load receptor.
5. Record the new indication as I2.
6. Calculate the difference in the two indications; I2 - I1.
7. Ensure that the change in indication determined in step 6 is greater than or equal to 0.7 times the extra load added in step 4.
8. Determine whether the instrument has passed or failed.
Digital Indicating Instrument
An additional load equal to 1.4 times the verification scale interval, when gently placed on or withdrawn from the instrument at equilibrium shall change the initial indication by one actual scale interval
[Clause 3(8)(ii)(b)
1. Zero the instrument.
2. Apply a load to the load receptor.
3. Apply additional standards weights of 0.1e until the indication changes up and stabilizes.
4. Record this indication.
5. Gently apply a load of 1.4e. The indication should increase by 1e to the next scale interval.
6. Determine whether the instrument has passed or failed.
7. Sensitivity
The sensitivity test is only performed on non-self-indicating instruments.
An extra load equivalent to the absolute value of the MPE for the applied load, shall be placed on the instrument at equilibrium and shall cause a permanent displacement of the indicating element should follow the "Indicator Displacement table, at least.
The sensitivity tests shall be carried out by placing extra loads with a slight impact, in order to eliminate the effects of discrimination threshold
1. Zero the instrument.
2. Apply a load to the load receptor.
3. Bring the instrument to its equilibrium position.
4. Apply an extra load equal to the absolute value of the MPE for the applied load.
5. Measure and record the linear distance between the middle points of this reading and the reading without the extra load as the permanent
displacement of the indication.
6. Determine whether the instrument has passed or failed.
8. Accuracy of Tare setting
This test is only required at first time verification only
A tare device shall permit setting the indication to zero with accuracy better than:
- ± 0.25e for electronic instruments and any instrument with analogue indication;
- ± 0.5d for mechanical instruments with digital indication and instruments with auxiliary indicating devices.
The operation of the tare device shall be visible on the indication panel with the sign "net".
For any tare load applied, the MPE for the remaining weighing capacity is the same as if no tare was being used
1. Load the instrument using a weight that exceeds the tare capacity and ensure the tare facility is not functional.
2. Activate the tare-setting device in the following way:
(a) Load the instrument using a weight that is within the tare-setting range.
(b) Add additional weights to take the total load just below the next changeover point.
3. Re-set the indication to zero using the tare-setting device.
4. Check the accuracy of tare-setting in the following way:
(a) Apply 10e to the load receptor.
(b) Apply an additional 0.25e. If the indication:
- remains unchanged go to step 4(c);
- changes up and stabilizes +1e from the original indication: FAIL
(c) If the indication remains unchanged in step 4(b), apply an additional 0.5e. If the indication:
- changes up and stabilizes +1e from the original indication: PASS
- remains unchanged: FAIL
(d) Remove 10e, 0.25e and 0.5e to bring the indication back to zero.
5. Check the tare weighing device in the following way:
(a) Ascertain whether the instrument has additive or subtractive tare.
(b) Determine the instruments remaining capacity.
(c) Add a load equal to full remaining capacity.
(d) Determine if the indication is within the MPE.
9. Price Computation
This test is only required at the initial verification
On a price-indicating instrument the supplementary primary indications are unit price and total price and, if applicable, number, unit price and total price for non-weighed articles, prices for non-weighed articles and price totals.
The price computation check is performed to check that the price calculating function is able to compute the total price. Conduct the check over a range of loads, and preferably during the weighing test.
1. Apply a load to the load receptor and input an appropriate unit price.
2. Calculate the total price from the unit price and indicated weight.
3. Compare the calculated total price with the indicated price.
4. Repeat the test at least four more times.
5. Determine whether the instrument has passed or failed.
View Model Test Report Part I | Model Test Report Part II
Sealing and Stamping
After all the testing is done and the instrument passes all tests, it is essential that the instrument be secured through sealing and stamping. Wilful or otherwise calibration of an instrument after a competent authority does the verification can change the metrological characteristic of the instrument without being visible. Sealing is mainly done to
- Prevent dismantling or maladjustment of components that might alter metrological characteristics of the instrument.
-
To control access to the calibration point of the instrument.
Using Paper Seal
The viable method is to put destructible self-adhesive type paper or thin plastic seal or sticker. Such seals, if attempted to remove, becomes defaced and any attempt of tempering can be easily detected.
The adhesive should be strong enough so that a seal cannot be removed without tearing it out. Such seals may carry
- A serial number, which may be noted in the Verification Certificate.
- All other marks put by punches now.
Mechanical Sealing
When a mechanical device like screw drivers can calibrate an instrument, then lead and wire seal or paper seal seals may be used to prevent access to this point.
Stamping
- The plate bearing the descriptive markings, fixed to the instrument, should be stamped.
- Stamping should be made on the soft metal plug provided on Dial Type Weigh-bridges
- On Weigh-bridge machines, other than dial type, the soft plug will be on the indicating lever or steel yard and this be stamped.
- In other machines, the soft plug should be on a conspicuous part of the beam or body.
- In all cases, the soft plugs should be made irremovable